Optimization of the ionization chamber structure for nuclear radiation measurement system

4 Nov 2025, 10:28
7m
60/Ground-106 - Lecture Hall (Administration Building)

60/Ground-106 - Lecture Hall

Administration Building

80
Show room on map
Student Competition Student Competition

Description

The modeled detector has been used to examine the versatility of the MCNP package in modeling the response of high pressure ion chambers to photon radiation. The results show that the response characteristics of a gas ion chamber filled with xenon are similar to the measured results from other studies of similar ion chambers. The model results indicate that while the use of a compressed xenon gas ion chamber detector improves the low energy response of the detector. Other detection gases like krypton, argon, neon, or helium can be easily modeled using the MCNP model created for this work, as was done for one part of this investigation.

Technical Track Student Competition

Primary author

Aya Samir (Fayoum University, Faculty of Science, Department of Physics)

Presentation materials